Defibtech 4 Year Lifeline View Battery
To check if your Defibtech View needs a new battery, look at the status indicator located next to the On/Off button. If it is not flashing green, press the centre “soft-key” button to activate the display screen. You will see the battery status indicator showing the amount of battery life remaining. The AED will also emit an audible “chirp” to let you know an alert condition exists. It is important to always leave the battery in the AED, keeping it in “Rescue Ready” mode.
Installation instructions:
- Remove the old battery by pushing the battery pack release latch located just below the battery and just above the electrodes storage slot. Wait 10 seconds before installing the new battery.
- Push the new battery into the slot until you hear the latch click. Battery is now securely installed.
- Once installed the Defibtech View will perform a battery pack insertion test. When completed, the unit will report the status of the battery pack, then return to standby mode.